The importance of patient positioning

In this post we discuss a case study detailing the importance of properly positioning and land marking the patient when performing a NeuroQuant® volumetric analysis.

glabellarWhen positioning a patient on the MRI table, land marking the X, Y and Z coordinates in the glabellar region is critical (see image). Land marking the patient in this region of the head, specifically the area between the eyebrows and above the nose, allows the center of the brain to align with the isocenter of the magnetic field. Magnetic field inhomogeneties are at their smallest magnitudes in isocenter. Therefore, positioning the head away from the isocenter in the + or – direction will cause greater distortion introduced by gradient nonlinearity.

If a sequence is acquired and uploaded to NeuroQuant with any of the X, Y or Z coordinates outside +/-100mm, automated volumetric analysis cannot be performed. An error message will be returned to the queue monitor and/or PACS. Below is an example of the NeuroQuant error message that would be generated in this case.

Unsuccessful patient positioning

Automated volumetric analysis is not possible in these instances. If NeuroQuant returns such an error message, re-scanning the patient with the scan origin centered appropriately is advisable.

Important note: If your patient is having an MRI performed on a body part other than the brain (i.e. cervical, thoracic, lumbar spine) it is very important to re-establish the glabellar region of the head as isocenter prior to acquiring the NeuroQuant volumetric scan.

Interested in learning more about how to ensure accurate NeuroQuant output results?

  • Explore the importance of using a 3D T1 sagittal MRI sequences for NeuroQuant analysis. You can read that blog post here.
  • Review how to ensure accurate NeuroQuant processing results. You can read that blog post here.
  • Understand why contrast agents are not used to achieve quality segmentation with NeuroQuant. You can read that blog post here.
  • Examine why good alignment to atlas is necessary. You can read that blog post here.

Additionally, more information about using and processing NeuroQuant can be found here.

References:

  1. Caramanos Z. et al. Gradient distortions in MRI: characterizing and correcting for their effects on SIENNA-generated measure of brain volume change. Neuroimage, 2010, Jan15:49(2):1601-11.
  1. Velthuizen R. et al. Review and evaluation of MRI nonuniformity for brain tumor response measurements. Med. phys. 25, 1655(1998).

Cortechs.ai | The importance of patient positioning

More Resources

05/29/2025

Enhancing Imaging Workflow with DICOM SR: NeuroQuant®’s Seamless Integration 

As imaging becomes more data-driven, tools like NeuroQuant® with DICOM SR support pave the way for smarter, more connected radiology

05/18/2025

Cortechs.ai and ALZ-NET Announce Strategic Collaboration to Improve Alzheimer’s Disease Care Through AI-Driven Imaging

This new strategic collaboration aims to enhance the quality of care for people living with Alzheimer’s disease.

05/16/2025

Case Study: NeuroQuant® for ARIA: Monitoring ARIA Progression in Early Alzheimer’s Disease

Dr. Suzie Bash highlights how advanced imaging with NeuroQuant for ARIA supported detection and monitoring of ARIA-E and ARIA-H in an 81-year-old patient.

05/13/2025

Dr. Ana M. Franceschi Joins Cortechs.ai Medical Advisory Board

Neuroimaging thought leader to enhance PET/MRI and dementia research impact through AI-powered tools

05/10/2025

All the Ways to Use the NeuroQuant Custom Report

With the ability to select structures from NeuroQuant’s 70+ structure parcellation atlas, the user has the power to truly personalize patient care!

05/04/2025

Introducing NeuroQuant MS Lesion Dynamics Computation: A Clearer Path to Understanding Disease Progression

Discover how the latest NeuroQuant MS update offers a clearer, more intuitive way to track lesion changes and understand disease progression over time.
Scroll to Top