Cortechs.ai | The importance of patient positioning

The importance of patient positioning

In this post we discuss a case study detailing the importance of properly positioning and land marking the patient when performing a NeuroQuant® volumetric analysis.

glabellarWhen positioning a patient on the MRI table, land marking the X, Y and Z coordinates in the glabellar region is critical (see image). Land marking the patient in this region of the head, specifically the area between the eyebrows and above the nose, allows the center of the brain to align with the isocenter of the magnetic field. Magnetic field inhomogeneties are at their smallest magnitudes in isocenter. Therefore, positioning the head away from the isocenter in the + or – direction will cause greater distortion introduced by gradient nonlinearity.

If a sequence is acquired and uploaded to NeuroQuant with any of the X, Y or Z coordinates outside +/-100mm, automated volumetric analysis cannot be performed. An error message will be returned to the queue monitor and/or PACS. Below is an example of the NeuroQuant error message that would be generated in this case.

Unsuccessful patient positioning

Automated volumetric analysis is not possible in these instances. If NeuroQuant returns such an error message, re-scanning the patient with the scan origin centered appropriately is advisable.

Important note: If your patient is having an MRI performed on a body part other than the brain (i.e. cervical, thoracic, lumbar spine) it is very important to re-establish the glabellar region of the head as isocenter prior to acquiring the NeuroQuant volumetric scan.

Interested in learning more about how to ensure accurate NeuroQuant output results?

  • Explore the importance of using a 3D T1 sagittal MRI sequences for NeuroQuant analysis. You can read that blog post here.
  • Review how to ensure accurate NeuroQuant processing results. You can read that blog post here.
  • Understand why contrast agents are not used to achieve quality segmentation with NeuroQuant. You can read that blog post here.
  • Examine why good alignment to atlas is necessary. You can read that blog post here.

Additionally, more information about using and processing NeuroQuant can be found here.

References:

  1. Caramanos Z. et al. Gradient distortions in MRI: characterizing and correcting for their effects on SIENNA-generated measure of brain volume change. Neuroimage, 2010, Jan15:49(2):1601-11.
  1. Velthuizen R. et al. Review and evaluation of MRI nonuniformity for brain tumor response measurements. Med. phys. 25, 1655(1998).

Cortechs.ai | The importance of patient positioning

More Resources

04/27/2026

Deep Learning Reconstruction and Quantitative MRI: Advancing Imaging Without Compromising Consistency

Deep learning reconstruction enhances MRI speed and clarity while preserving the consistency and reliability of quantitative imaging

04/10/2026

OnQ Prostate Vendor Validation: From Research to Clinical Adoption

OnQ Prostate delivers consistent, reliable AI-powered prostate MRI insights across vendors, ensuring confident diagnosis and streamlined clinical workflows.

03/20/2026

Quantitative MRI, the McDonald Criteria, and the Future of Multiple Sclerosis Care

MRI is the foundation for diagnosing and monitoring multiple sclerosis (MS)

03/13/2026

Cortechs Cares: Community Impact and Volunteer Efforts in 2025

Cortechs Cares highlights how our team gave back in 2025 through community volunteer efforts, including an Alzheimer’s Walk and holiday toy drive.

03/11/2026

OnQ™ Prostate: Bringing Quantitative MRI to Prostate Imaging

Advancing prostate MRI with quantitative diffusion biomarkers. OnQ™ Prostate brings objective RSI-based insights to support data-driven cancer evaluation.

03/04/2026

Cortechs.ai and Siemens Healthineers Partner to Expand Global Access to NeuroQuant® Technologies

NeuroQuant® Lesion Surveillance will be integrated into the Siemens Healthineers Digital Marketplace to advance quantitative brain imaging worldwide.
Scroll to Top