Cortechs.ai | Are you positioning patients correctly?

Are you positioning patients correctly?

In this post we discuss a case study detailing the importance of properly positioning and land marking the patient when performing a NeuroQuant® volumetric analysis.

glabellarWhen positioning a patient on the MRI table, land marking the X, Y and Z coordinates in the glabellar region is critical (see image). Land marking the patient in this region of the head, specifically the area between the eyebrows and above the nose, allows the center of the brain to align with the isocenter of the magnetic field. Magnetic field inhomogeneties are at their smallest magnitudes in isocenter. Therefore, positioning the head away from the isocenter in the + or – direction will cause greater distortion introduced by gradient nonlinearity.

If a sequence is acquired and uploaded to NeuroQuant with any of the X, Y or Z coordinates outside +/-100mm, automated volumetric analysis cannot be performed. An error message will be returned to the queue monitor and/or PACS. Below is an example of the NeuroQuant error message that would be generated in this case.

Cortechs.ai | Are you positioning patients correctly?

Automated volumetric analysis is not possible in these instances. If NeuroQuant returns such an error message, re-scanning the patient with the scan origin centered appropriately is advisable.

Important note: If your patient is having an MRI performed on a body part other than the brain (i.e. cervical, thoracic, lumbar spine) it is very important to re-establish the glabellar region of the head as isocenter prior to acquiring the NeuroQuant volumetric scan.

Interested in learning more about how to ensure accurate NeuroQuant output results?

  • Explore the importance of using a 3D T1 sagittal MRI sequences for NeuroQuant analysis. You can read that blog post here.
  • Review how to ensure accurate NeuroQuant processing results. You can read that blog post here.
  • Understand why contrast agents are not used to achieve quality segmentation with NeuroQuant. You can read that blog post here.
  • Examine why good alignment to atlas is necessary. You can read that blog post here.

Additionally, more information about using and processing NeuroQuant can be found here.

References:

  1. Caramanos Z. et al. Gradient distortions in MRI: characterizing and correcting for their effects on SIENNA-generated measure of brain volume change. Neuroimage, 2010, Jan15:49(2):1601-11.
  1. Velthuizen R. et al. Review and evaluation of MRI nonuniformity for brain tumor response measurements. Med. phys. 25, 1655(1998).

More Resources

02/25/2026

Cortechs.ai Announces MFDS (KFDA) Approval of Global-Leading NeuroQuant® V5 Platform in South Korea

MFDS approves NeuroQuant® V5 in South Korea, expanding AI-driven volumetric neuroimaging and clinical decision support.

02/24/2026

Financial Conflict Of Interest Policy Applicable To NIH Awards

02/20/2026

NeuroQuant® PET: FDA 510(k)-Pending Advancement in Molecular Dementia Imaging

FDA 510(k)-Pending NeuroQuant® PET delivers automated, standardized quantification for amyloid, tau, and FDG imaging in dementia care.

02/19/2026

How Cortechs.ai Protects ePHI: HIPAA‑Aligned Controls—and Beyond

Cortechs.ai’s HIPAA-aligned, defense-in-depth security safeguards ePHI across AI workflows—exceeding baseline compliance standards.

02/11/2026

NeuroQuant Normative Database: Advancing Age‑ and Sex‑Specific Brain Volumetry

Learn how the latest updates to the NeuroQuant normative database white paper strengthen age- and sex-specific brain volumetry and enhance clinical confidence.

02/06/2026

NeuroQuant Normative Database: A Standard for Age- and Sex-Specific Brain Volumetry

Whitepaper: How Cortechs.ai’s normative database was created to accurately compare an individual’s brain structure volume to a healthy population.
Scroll to Top